MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION VISTA 500; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION VISTA 500; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number DIMENSION VISTA 500
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/14/2017
Event Type  malfunction  
Manufacturer Narrative
The customer contacted a siemens customer care center (ccc) specialist.The ccc reviewed the data supplied.The customer's quality control (qc) was found to be out of range at the time of the event.A siemens customer service engineer (cse) was dispatched to the customer's site.The cse performed all challenge methods and replaced the reagent.The cause of the discordant, falsely elevated hemoglobin a1c results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Event Description
Discordant, falsely elevated hemoglobin a1c results were obtained on a patient sample on a dimension vista 500 instrument.The initial result was released to the physician(s).The customer repeated the same sample on the same dimension vista instrument, resulting lower.The customer issued a corrected report to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated hemoglobin a1c results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIMENSION VISTA 500
Type of Device
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive
po box 6101
newark DE 19714 6101
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC
registration number: 1226181
101 silvermine road
brookfield CT 06804
Manufacturer Contact
timothy rice
511 benedict ave
tarrytown, NY 10591
9145242406
MDR Report Key6559081
MDR Text Key74910413
Report Number2517506-2017-00483
Device Sequence Number0
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051087
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2017
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDIMENSION VISTA 500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/15/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/11/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-