• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 40 PARKER RAD-8; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER RAD-8; OXIMETER Back to Search Results
Model Number 22042
Device Problem No Audible Alarm (1019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/20/2017
Event Type  malfunction  
Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event Description
Customer reported "reads 'sensor off' but doesn't alarm." no known impact or consequence to patient was reported.
 
Manufacturer Narrative
The returned device was evaluated.During inspection, the unit was unable to power on via battery power.Using ac power, the device was able to obtain readings and alarmed audibly and visually under alarm conditions.The brightness setting key, switch 5, was non responsive on the keypad and showed signs of physical damage.As a result, the alarm functions were inaccessible.Internal inspection showed the battery was disconnected.The customer complaint could not be duplicated.A service history record review reveals that this unit was in the field for over two (2) years with no previous reported issues related to this reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-8
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
Manufacturer (Section G)
INDUSTRIAL VALLERA DE MEXICALI S.A DE C.V.
calzada del oro, no. 2001
parque industrial palaco
mexicali, baja california 21600
MX   21600
Manufacturer Contact
stefan lissmann
52 discovery
irvine, CA 92618-1604
9492977000
MDR Report Key6565011
MDR Text Key75099136
Report Number2031172-2017-00552
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092838
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number22042
Device Catalogue Number9190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received07/17/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-