Product analysis: upon receipt at medtronic¿s quality laboratory, the returned product specimen was visually examined.All leaflets were in the closed position.Two cuts through the tunica of the right cusp were observed.The 4mm and 5mm cuts appeared consistent with a puncture or laceration by a sharp instrument such as forceps or scalpel.A tear between the cuts was observed.All commissures were intact.The right noncoronary commissure (nrc) post was relaxed due to a broken suture.The broken suture tips were found tucked in the nrc post and within the sewing ring adjacent to the right cusp (rc).Under magnification, the tips of the sutures were observed.Both suture tips were frayed/jagged not smooth, indicating they were broken rather than cut.A small smooth indentation was observed on one of the suture tips.Its smooth appearance suggests damaged due to a sharp object.As received, the valve holder was activated as the left right commissure (lrc) and noncoronary left commissure (nlc) posts were deflected.Scrape like damage and indentations were observed on top of the holder, possibly caused by sharp object such as forceps.The same scrape-like damage and indentations was noted on both sides of the holder leg adjacent to the rc.The device was inspected by a manufacturing engineer during product analysis.It was confirmed that these observed damages on the leaflet, suture and holder could not have been made during the manufacture of the device and its components.Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.The observations of the cuts were appeared to be consistent with a puncture or laceration by a sharp instrument such as forceps during implant.Per medtronic procedure each hancock ii valve was 100 percent inspected during final inspection on tissue characteristics and damages.In addition, the valve holder suture was broken and the valve holder was damaged; these additional observations also appeared to be consistent with damage by a sharp instrument such as forceps during implant or explant (removed the holder from the handle).Based on the received information and analysis results of the returned device, the clinical observations, tear, was confirmed.The cause of the tear could be related to the implant procedure contact with a sharp instrument.If information is provided in the future, a supplemental report will be issued.
|