A patient was treated in therapeutic plasma exchange (tpe), using a prismaflex tpe 2000 set and a prismaflex control unit.There was reportedly no issue during the priming of the set, but the customer noted a leakage from the effluent line of the set after treatment start.They reportedly noticed a hole in the line, causing the leakage.Treatment was discontinued and the extracorporeal blood circuit was returned to the patient.No medical intervention was needed due to the event.The event reportedly resulted in the loss of 100 ml of albumin for the patient.No additional information is available.
|