Model Number M00545000
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/12/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation results: visual evaluation of the returned device found that residues were present which indicate use and handling.Further evaluation noted that the working length including the catheter and pull wire had been kinked in multiple locations throughout.The pull wire was found broken at the distal end and the bristled portion of the brush was not present and was not returned for evaluation.In addition, the pull wire presents drag marks that were probably caused by the customer at the time the pull wire was cut off.The guidewire channel was torn all the way until the distal tip and the ring was not present and was not returned for analysis.Moreover, the extrusion portion had been torn where the ro marker was supposed to be.Functional analysis was not performed due to the device condition.The reported complaint was not confirmed.Due to anatomical/procedural factors encountered during the procedure; performance of the device was limited.Therefore, the most probable root cause classification is operational context.A review of the device history record (dhr) was performed and no deviations were found.
 
Event Description
It was reported to boston scientific corporation that an rx cytology brush was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2017.According to the complainant, during the procedure, the brush failed to extend and retract.The procedure was completed with this device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.This event has been deemed a reportable event based on the investigation results; ro marker detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX CYTOLOGY BRUSH
Type of Device
ENDOSCOPIC CYTOLOGY BRUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer (Section G)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6571250
MDR Text Key75322666
Report Number3005099803-2017-01394
Device Sequence Number1
Product Code FDX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K930348
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/16/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date10/31/2018
Device Model NumberM00545000
Device Catalogue Number4500
Device Lot Number19891874
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/14/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/27/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-