U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS FOOTED ATTACHMENT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS FOOTED ATTACHMENT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number 5407-FA2-000
Device Problems Charred (1086); Overheating of Device (1437); Device Operates Differently Than Expected (2913)
Patient Problem No Information (3190)
Event Date 04/24/2017
Event Type  malfunction  
Event Description
Bearings failed, causing drill bit to run on surface of guard, therefore leading to heat and charring.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOTED ATTACHMENT
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
STRYKER INSTRUMENTS
4100 east milham ave.
kalamazoo MI 49001
MDR Report Key6571945
MDR Text Key75332325
Report Number6571945
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 05/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number5407-FA2-000
Device Catalogue Number5407-FA2-000
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA05/02/2017
Event Location Hospital
Date Report to Manufacturer05/02/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age46 YR
-
-