• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FRONTIER II; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FRONTIER II; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number 5586
Device Problem No Device Output (1435)
Patient Problem Loss Of Pulse (2562)
Event Date 04/03/2017
Event Type  Injury  
Event Description
It was reported that patient presented to the hospital for a generator change due to normal eri.During the procedure, the device exhibited temporarily loss of output, as a result of the cautery exposure.The patient experienced asystole, and the patient was connected to a temporary pacing system.The procedure continued without complication.The patient was stable and in good condition prior to and post procedure.
 
Manufacturer Narrative
The damage found was sustained during the surgical procedure.The device was otherwise normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FRONTIER II
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6574329
MDR Text Key75399685
Report Number2017865-2017-02695
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2008
Device Model Number5586
Device Lot Number0002134181
Other Device ID Number05414734502948
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/15/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age82 YR
-
-