A siemens customer service engineer (cse) was dispatched to the customer site.After analyzing the instrument, the cse removed the integrated multi-sensor technology (imt) module and replaced it with a new one.The cse replaced the imt rotary valve and motor.The cse power flushed the imt system and realigned it.The cse replaced the imt probe and realigned it.The cse performed quick check, which passed and ran imt quality control (qc), resulting within acceptable ranges.The cse ran precision and patient comparison testing, resulting satisfactory.A regional support center (rsc) specialist reviewed the instrument data with the technical applications specialist (tas) which indicated poor sample handling.The rsc specialist instructed the tas to discuss with the customer the use of proper sample handling techniques to minimize potential sample integrity issues.The cause of the discordant, falsely elevated na results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|
Discordant, falsely elevated sodium (na) results were obtained on patient samples on a dimension vista 500 instrument.The discordant results were reported to the physician(s), who questioned them.New samples obtained from the patients were tested on an alternate dimension vista instrument, resulting lower.The new sample results were reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated na results.
|