On (b)(6) 2017 customer using a tosoh g8 reported that two patient hba1c results from the same run were sent to the host lis that were incorrect and the results were then reported to the physician.One result was questioned by the physician: patient 1 hba1c = 1.3%, patient 2 hba1c = 3.2%.On (b)(6) 2017 tosoh bioscience notified of the incorrect patient results being released to the physician.Customer indicated that the g8 column count = 7059 (which is over the recommended 2500 column count).Hba1c qc results were within target range.When tosoh bioscience technical support specialist reviewed the chromatograms for the reported results the chromatograms did not meet parameters for acceptability.The chromatograms had elevated baselines and multiple missing peaks which indicated that the results were not reportable but had been released without having the required chromatogram review by the technologist.The expired column was replaced with a new column, qcs and patient specimens were repeated.Qcs and patient specimen chromatograms had all peaks present, correct baseline re-established, and correct system pressure maintained per new column inspection report.Root cause: g8 column count exceeded recommended column count.
|