Model Number CNS-6201A
Device Problems Failure to Power Up (1476); Device Inoperable (1663)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 04/27/2017
Event Type  malfunction  
Manufacturer Narrative
The customer reported that they heard a pop sound from the central nurses station (cns) and the system crashed and will not power on.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The customer reported that they heard a pop sound from the central nurses station (cns) and the system crashed and will not power on.
 
Manufacturer Narrative
The customer reported that they heard a pop sound from the central nurses station (cns) and the system crashed and will not power on.The unit was cleaned and evaluated.The reported problem of system not powering up was duplicated, and our evaluation also found that the unit has two hdd malfunctioning.All malfunctioning parts were replaced.The unit was tested per the operator's/service manual and the unit completed extended testing and operates to manufacturer's specifications.Nihon kohden will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CNS-6201A
Type of Device
CENTRAL MONITOR SYSTEM
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, japan 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgm
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
2687488
MDR Report Key6586621
MDR Text Key75920039
Report Number8030229-2017-00150
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04931921114131
UDI-Public4931921114131
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102376
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type study,user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/23/2017,04/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCNS-6201A
Device Catalogue NumberPU-621RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA05/23/2017
Distributor Facility Aware Date04/27/2017
Device Age13 MO
Event Location Hospital
Date Report to Manufacturer05/23/2017
Date Manufacturer Received05/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-