Catalog Number PXSLIMLAN135T45 |
Device Problems
Kinked (1339); Device Handling Problem (3265)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/27/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
|
Event Description
|
The patient was undergoing a coil embolization procedure using a lantern delivery microcatheters (lantern).During the procedure, the physician successfully deployed and detached three ruby coils in the target vessel using a lantern.The physician then inadvertently kinked the lantern near the hub and noticed it when trying to flush the lantern with saline.Therefore, the lantern was removed and the procedure was successfully completed using a new lantern.There was no report of an adverse effect to the patient.
|
|
Manufacturer Narrative
|
The lantern delivery microcatheters (lantern) was fractured approximately 7.0 cm from the proximal hub.The distal tip of the microcatheter had multiple bends at its end.Evaluation of the returned lantern revealed that the microcatheter was fractured near the proximal hub.If the microcatheter is forcefully manipulated during use, this type of damage is likely to occur.Further evaluation revealed multiple bends on the distal tip of the catheter.This damage is likely from the lantern remaining in tortuous anatomy during the procedure.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Search Alerts/Recalls
|