Miltenyi biotec inc.Initiated a maintenance service on 16th may 2017 which included testing the pump unit movement and tolerances.The results were laid dov/n in service report id #(b)(4).There were no deficiencies found at the time of the service.Furthermore, the cam team of miltenyi biotec inc.Also noted that they provided the customer with an extra tubing set, buffer, reagent and frozen cell sample.As a result, the customer was able to run the mock sample without any errors.Therefore, the complained failure couldn't be confirmed by miltenyi biotec.
|