• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012450-08
Device Problems Inflation Problem (1310); Nonstandard Device (1420)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/03/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.The complaint investigation determined the reported difficulty was related to manufacturing issues associated with the protective sheath.On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints.To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda on march 17, 2017 [medwatch # 2024168-2017-02310].Corrective action has been implemented per site operating procedures.The product will continue to be trended.(b)(4).
 
Event Description
Reportedly, a 3.25x08mm nc trek balloon dilatation catheter failed to inflate.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6593982
MDR Text Key76186520
Report Number2024168-2017-04540
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08717648151910
UDI-Public(01)08717648151910(17)190430(10)60530G1
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K110134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial
Report Date 05/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2019
Device Catalogue Number1012450-08
Device Lot Number60530G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/04/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2037-2039-2017
Patient Sequence Number1
-
-