• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-523LNAP
Device Problems Break (1069); Positioning Problem (3009)
Patient Problem Hyperglycemia (1905)
Event Date 04/28/2017
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump's reservoir compartment was damaged and the reservoir was no longer secure.Blood glucose level at the time of the incident was 250 mg/dl.The customer was advised that the insulin pump would be replaced and agreed to return the product for analysis.
 
Manufacturer Narrative
The insulin pump was received with completely broken off reservoir tube lip and test reservoir will not lock or click into place.Unable to perform the basic occlusion test and occlusion test due to broken reservoir tube lip.The insulin pump passed the operating currents measurement, self-test, unexpected restart error test, displacement test, prime and excessive no delivery test.The insulin pump had cracked reservoir tube and minor scratched display window.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6595049
MDR Text Key76191440
Report Number3004209178-2017-70457
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 06/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-523LNAP
Device Catalogue NumberMMT-523LNAP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received06/27/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age21 YR
Patient Weight47
-
-