Catalog Number 050-12E |
Device Problems
Insufficient Heating (1287); Temperature Problem (3022)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/05/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).A visual, dimensional, and functional inspection of the device involved in the complaint could not be conducted since the device was not returned at the time of this report.A device history record investigation did not show issues related to this complaint.A record assessment (fmea) was conducted and no changes required.Corrective actions cannot be established since it is necessary to receive the physical sample to perform a proper investigation and confirm the alleged defect.Customer complaint cannot be confirmed based on the information received.If the device sample becomes available at a later date, this complaint will be updated accordingly.
|
|
Event Description
|
Customer complaint alleges that the doctor in the "icu dept, (b)(6) hospital, found the nebulizer heater can't be heated during clinic use".It was reported there was no medical intervention necessary.Patient condition reported as "fine".
|
|
Manufacturer Narrative
|
(b)(4).The sample was returned for evaluation.A visual exam was performed and no defects were observed.Functional testing was also conducted and the unit passed the tests.It heated up on minimum (50.8 c), medium (86.7 c), and maximum (126.9 c) temperature settings during the testing and no issues were encountered.Based on the investigation performed, the reported complaint could not be confirmed.There were no issues found with the returned device.
|
|
Event Description
|
Customer complaint alleges that the doctor in the "icu dept, (b)(6), found the nebulizer heater can't be heated during clinic use".It was reported there was no medical intervention necessary.Patient condition reported as "fine".
|
|