• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON AQUATHERM III,UNITED KINGDOM; BREATHING SYSTEM HEATER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON AQUATHERM III,UNITED KINGDOM; BREATHING SYSTEM HEATER Back to Search Results
Catalog Number 050-12E
Device Problems Insufficient Heating (1287); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/05/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, dimensional, and functional inspection of the device involved in the complaint could not be conducted since the device was not returned at the time of this report.A device history record investigation did not show issues related to this complaint.A record assessment (fmea) was conducted and no changes required.Corrective actions cannot be established since it is necessary to receive the physical sample to perform a proper investigation and confirm the alleged defect.Customer complaint cannot be confirmed based on the information received.If the device sample becomes available at a later date, this complaint will be updated accordingly.
 
Event Description
Customer complaint alleges that the doctor in the "icu dept, (b)(6) hospital, found the nebulizer heater can't be heated during clinic use".It was reported there was no medical intervention necessary.Patient condition reported as "fine".
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and no defects were observed.Functional testing was also conducted and the unit passed the tests.It heated up on minimum (50.8 c), medium (86.7 c), and maximum (126.9 c) temperature settings during the testing and no issues were encountered.Based on the investigation performed, the reported complaint could not be confirmed.There were no issues found with the returned device.
 
Event Description
Customer complaint alleges that the doctor in the "icu dept, (b)(6), found the nebulizer heater can't be heated during clinic use".It was reported there was no medical intervention necessary.Patient condition reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON AQUATHERM III,UNITED KINGDOM
Type of Device
BREATHING SYSTEM HEATER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key6597843
MDR Text Key76311917
Report Number3003898360-2017-00670
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/30/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number050-12E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received08/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-