The hydromark breast biopsy site marker is used to mark tissue during a percutaneous breast biopsy procedure, visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and mri.One 4010-01-08-t4 hydromark was received on 5/30/2017 and investigated on 6/2/2017.The device was received in good condition.The device shows signs of being used in a procedure.The fixed ramp is missing from the device and was not returned thus preventing a full investigation or determining root cause.The device history records were reviewed and there were no non-conformances that would relate to this failure.It was confirmed that the pink plastic pledget (fixed ramp) was found on the tip of the probe and not left in the biopsy site.However, as the occurrence has the potential to result in patient consequence with the need for surgical intervention or other form of treatment, and pursuant to 21 cfr 803, we are submitting this medwatch report.
|