• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723NAS
Device Problems Device Displays Incorrect Message (2591); Positioning Problem (3009)
Patient Problem Hypoglycemia (1912)
Event Date 05/06/2017
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump had a motor error alarm and multiple motor position encoder error alarms.Blood glucose level at the time of the incident was 65 mg/dl.Low blood glucose troubleshooting was declined.Customer stated that the device may have been exposed to high magnetic fields.The customer was advised that the insulin pump would be replaced and agreed to return the device for analysis.
 
Manufacturer Narrative
Motor error alarm during rewind due to motor encoder signal out of phase.Unable to perform self test, off no power test, unexpected restart error test, occlusion test, prime/test, excessive no delivery test, displacement test due to motor error alarm.Pump passed idle current test and run current test.Unable to verify motor position encoder error alarm due to motor error alarm during rewind.Pump had cracked reservoir tube lip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAS
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key6614339
MDR Text Key76947653
Report Number3004209178-2017-52495
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169507258
UDI-Public(01)00643169507258
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 07/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723NAS
Device Catalogue NumberMMT-723NAS
Device Lot NumberB5723NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received06/23/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age23 YR
Patient Weight160
-
-