Device evaluated by manufacturer: examination of the returned complaint device revealed that the returned product consisted of the watchman access sheath (was).The valve was opened as received and blood was on the device.The valve, hub, shaft, and tip were examined.There were multiple kinks found along the shaft of the device.The tip was damaged with shaft material folded back onto the device.Functional testing of the valve confirmed the ability to completely close the valve with minimal forward force.Water was injected into the was by attaching a syringe filled with water.There were no leaks or air bubbles observed during water injection.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|