MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754CAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-754CAB |
Device Problems
Device Displays Incorrect Message (2591); Positioning Problem (3009)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/01/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned. no conclusion can be drawn at this time. we therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states. however, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
The customer reported via phone call the insulin pump alarmed motor error and a motor position encoder error. the customer¿s blood glucose level was 49 mg/dl at the time of the incident. the customer was advised to change the infusion set and call back. the insulin pump will not be returned for analysis.
|
|
Manufacturer Narrative
|
Motor error alarm during the basic occlusion test due to faulty force sensor resistor.The motor was tested outside of the device and passed.The insulin pump passed displacement test, self test and unexpected restart error test.The insulin pump passed all operating currents tested within the spec range and passed off no power test.Pump received with cracked reservoir tube lip, cracked case near display window corners and minor scratches on display window noted.
|
|