The customer complained of multiple sample up/down alarms and an erroneous low result for 1 patient tested for phos2 phosphate (inorganic) ver.2 (phos2) on a cobas 6000 c (501) module.The initial phos2 result was 0.1 mg/dl with a data flag.The lab tech was not paying close attention and reported the result outside of the laboratory.The sample was repeated on a different analyzer and the result was 4.2 mg/dl.The repeat result was believed to be correct and a corrected report was issued.The patient was treated with phosphorus based on the initial phos2 result reported outside of the laboratory.There was no allegation that an adverse event occurred due to this treatment.The customer stated there were other erroneous results reported outside of the laboratory.The customer did not have the specific results, however, the repeat results were not deemed to be clinically significant when compared to the results initially reported outside of the laboratory.The only clinically significant result the customer had was the low, erroneous phos2 result.The phos2 reagent lot number was 19635201 with an expiration date of 01/31/2018.The customer checked the sample probe and found that there was gel in it.The customer does not have another probe so she will clean the probe.The field service engineer visited the customer site after the customer had cleaned the sample probe.The fse ran a pipetting accuracy check.The customer ran quality controls.The results from all instrument tests were acceptable.
|
Medical assessment of the event was performed and noted that the initial phos2 result was accompanied by a data flag and should not have been reported outside of the laboratory.The decision to treat the patient after one questionable, plausible result could not be addressed as the medical background of the patient is not known (e.G.Kidney disease, patient in intensive care unit).A product problem was not identified.
|