Model Number MMT-1712W
Device Problems Device Displays Incorrect Message (2591); Naturally Worn (2988); Scratched Material (3020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/13/2017
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
Customer reported via phone call replace battery now alarm on the insulin pump.Customer¿s blood glucose level was 233 mg/dl.Troubleshooting for the alarm was performed.Customer advised they replaced the battery on the device multiple times and the issue remained unresolved.Customer received the replace battery now alarm for the second time.Customer advised the display screen had several scratches on it and the buttons were worn out.Customer was advised to discontinue use of the device and revert to a backup plan.Customer was advised that the device would be replaced and agreed to return the product for analysis.
 
Manufacturer Narrative
Insulin pump passed sleep current measurement, active current measurement, self test and displacement test.No unexpected replace battery now alarm or power loss alarm noted during testing.Insulin pump was received with minor scratched lcd window, scratched case and cracked select button keypad overlay.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
640G INSULIN PUMP MMT-1712W
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key6623614
MDR Text Key77297601
Report Number3004209178-2017-53901
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169577701
UDI-Public(01)00643169577701
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1712W
Device Catalogue NumberMMT-1712W
Device Lot NumberHG0QTH1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2018
Date Manufacturer Received08/28/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age11 YR
-
-