Model Number 694765 |
Device Problems
Mechanical Problem (1384); Cut In Material (2454); Decreased Sensitivity (2534); Connection Problem (2900)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/15/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
This device was reported as included in the field action.Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action.It is included in the field action in the abundance of caution.
|
|
Event Description
|
It was reported that the right ventricular (rv) lead exhibited chronic low r-waves.The lead was capped and replaced.During the change-out procedure, the superior vena cava (svc) pin of the right ventricular (rv) lead would not come out of the implantable cardioverter defibrillator (icd) header.Three different wrenches were used but the setscrew was unable to be engaged.The physician observed a lot of material in the grommet and had to cut the svc pin off.The svc pin was left within the header of the device and the icd was explanted and replaced.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|