The customer received questionable ca2 calcium gen.2 results for two patient samples.Of the data provided, only the erroneous result for one patient sample was reported outside the laboratory.The initial result was 5.7 mg/dl and was reported outside the laboratory.The doctor sent the patient to the emergency room (er) and a new sample was drawn with a result of 10.1 mg/dl.The original sample was repeated on (b)(6) 2017 and the result was 9.9 mg/dl.The repeat result was considered correct.The sample was also tested on the integra 400 and the result matched the repeat result.No specific data from the integra 400 was provided.The customer did not know of any treatment provided based on the erroneous result and the patient was discharged from er.No adverse event was reported.The reagent lot number was 22326401 with an expiration date of 5/31/2018.The field service representative found an issue with the sample probe.He adjusted the internal sample probe pressures, replaced the vacuum line on the rinse mechanism, adjusted the gear pump head, and checked the pressure.The customer ran qc which passed within their specifications.The field application specialist did not find an issue with the application.He checked the application settings, had the customer add their critical range repeat values, and verified calibration and qc had been acceptable.He verified no other assays showed performance problems.Upon follow up, the customer confirmed there were no further issues after the service visit.Based on review of the provided calibration and qc data, a general reagent issue was excluded.
|