The customer obtained a questionable low test result for one patient sample using the elecsys ferritin assay on the cobas 8000 e 602 module (e602).The initial ferritin result was 0.588 ng/ml.The repeat result was 3105 ng/ml.No adverse event occurred.The ferritin reagent lot number and expiration date was requested but not provided.Calibration was successful prior to the event.Customer performed a reagent probe check and the analyzer generated an abnormal reagent probe movement alarm.He believed the probe was hitting a closed reagent lid.The field service representative found the sample pipetter liquid level detector (lld) circuit was out of adjustment.He adjusted the lld, performed an instrument overview, and performed mechanism checks which passed.He conducted an assay performance check which was acceptable.The customer performed calibration and qc afterwards which passed.The instrument performed within specifications.The root cause for this event was the maladjusted sample pipetter lld circuit.This can cause an early lld signal, and therefore too little sample is pipetted.After service was performed, the analyzer worked according to specifications.
|