• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500 BLOOD GAS ANALYZER; RP 500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500 BLOOD GAS ANALYZER; RP 500 Back to Search Results
Catalog Number 10492730
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/16/2017
Event Type  malfunction  
Manufacturer Narrative
Siemens has reviewed the data files provided by the customer and come to the following conclusion: for sample "a," due to the repeat samples at 1:44 and 2:06 on different rp500s using different draws having the same result, it is likely that the 1:32 sample was exposed to pre-analytical po2 contamination.For sample "b," it is likely that the 03:19 sample was exposed to pre-analytical po2 contamination in the 15 minutes between samples.
 
Event Description
The customer reported discrepant po2 results on two different rp 500 analyzers.There was no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500 BLOOD GAS ANALYZER
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key6633370
MDR Text Key77518172
Report Number3002637618-2017-00094
Device Sequence Number1
Product Code CHL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Respiratory Therapist
Type of Report Initial
Report Date 06/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Respiratory Therapist
Device Catalogue Number10492730
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/16/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-