MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP Back to Search Results
Model Number 801041
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/19/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).As per field service representative (fsr), the system 1 tube clamps have a knob for opening each side of the clamp.The customer opened it up and found a broken piece of plastic.The fsr sent a replacement tube clamp to the customer.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the roller pump tube clamp was stuck and the knob would not turn.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Evaluation is in progress, but not yet concluded.
 
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) could not rotate the tube clamp knob.Internal inspection revealed a broken cam on left lower slide affecting proper operation of the tube clamp assembly.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6634060
MDR Text Key77543220
Report Number1828100-2017-00264
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131618
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup,Followup
Report Date 09/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801041
Device Catalogue Number801041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/16/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/23/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number1828100-2/03/2011-003-C
Patient Sequence Number1
-
-