Model Number 801041 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/19/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).As per field service representative (fsr), the system 1 tube clamps have a knob for opening each side of the clamp.The customer opened it up and found a broken piece of plastic.The fsr sent a replacement tube clamp to the customer.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the roller pump tube clamp was stuck and the knob would not turn.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Evaluation is in progress, but not yet concluded.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician (pst) could not rotate the tube clamp knob.Internal inspection revealed a broken cam on left lower slide affecting proper operation of the tube clamp assembly.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|