• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE XVG® THROMBECTOMY SET; CATHETER, EMBOLECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE XVG® THROMBECTOMY SET; CATHETER, EMBOLECTOMY Back to Search Results
Model Number 105042-001
Device Problems Device Displays Incorrect Message (2591); Aspiration Issue (2883)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/23/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr.: the device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined.(b)(4).
 
Event Description
It was reported that an error message displayed during aspiration.An xvg® thrombectomy set was selected for a thrombectomy procedure in the distal leg artery.The device primed fine and was introduced into the patient's body.After about 5 seconds of aspiration, an error message displayed and it was observed the pump bulb was leaking saline.The procedure was completed with a different bigger angiojet catheter.There were no patient complications and the patient was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XVG® THROMBECTOMY SET
Type of Device
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6636246
MDR Text Key77534880
Report Number2134265-2017-05839
Device Sequence Number1
Product Code DXE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K072769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/13/2017
Device Model Number105042-001
Device Lot Number18960833
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/23/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-