• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG PLAINFIELD Kangaroo; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG PLAINFIELD Kangaroo; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 383401
Device Problem Device Stops Intermittently (1599)
Patient Problem No Patient Involvement (2645)
Event Date 05/17/2017
Event Type  malfunction  
Event Description
According to the reporter, the pump was returned because it shuts off and on during testing.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed for the reported condition of intermittent power.The unit was triaged and the reported issue could not be confirmed at this time.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Kangaroo
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN MFG PLAINFIELD
2824 airwest blvd
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN MFG PLAINFIELD
2824 airwest blvd
plainfield IN 46168
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key6636669
MDR Text Key77616754
Report Number3008361498-2017-05186
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number383401
Device Catalogue Number383401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received07/05/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-