• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM; BALLOON EXPANDABLE VASCULAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM; BALLOON EXPANDABLE VASCULAR COVERED STENT Back to Search Results
Catalog Number LSM1350938
Device Problem Burst Container or Vessel (1074)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/03/2017
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Receipt of device is pending.The investigation is currently in progress.Not returned.
 
Event Description
It was reported that the balloon burst during the procedure.Another stent of a longer size was used to complete the procedure.There was no patient injury reported.
 
Manufacturer Narrative
Updating for new information ¿ the device has been received and based on the evaluation of the device this event is no longer determined to be reportable.This is based on the findings from the evaluation where a pinhole was identified in the balloon.A pinhole rupture is not known to have caused or contributed to any injury and there is no history of this type of failure being reported.Therefore this complaint is now determined to be not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTREAM
Type of Device
BALLOON EXPANDABLE VASCULAR COVERED STENT
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford
EI 
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford
EI  
Manufacturer Contact
fiona nimhullain
539237111
MDR Report Key6639901
MDR Text Key77835721
Report Number9616666-2017-00027
Device Sequence Number1
Product Code PRL
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
P160024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2019
Device Catalogue NumberLSM1350938
Device Lot NumberCMAU0029
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2017
Is the Reporter a Health Professional? Yes
Event Location Hospital
Date Manufacturer Received06/17/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/05/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-