It was reported that during set-up of the device for a cardiopulmonary bypass procedure (cpb), the ring around the top of the saw where the blade goes in was missing.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.Per clinical review on june 5, 2017: the sternal saw issue occured during set-up.The saw system is tested prior to the patient arriving in the operating room.During test, the saw would run, then stop, then run, and stop.It did not continuously operate.The saw was not used for the procedure and a back-up was used.
|