The customer contacted the siemens customer care center (ccc) to report the discordant calcium_2 (ca2_c) result, which occurred after water maintenance on the system.The customer provided quality control (qc) data, which indicated results were within range on the day of event.The ccc evaluated the data file, and stated the cause of the issue was the water source.A siemens customer service engineer (cse) was dispatched to the customer site.The cse ran couple of washes.The cse primed the instrument, and flushed water through the system to remove excess dissolved solid.The cse ran qc, and was in range.The cause of the discordant falsely low calcium_2 (ca2_c) result is due to source water.The instrument is performing according to the specifications.No further evaluation of this device is required.
|