Sedecal integrated console display out.Patient was on the table and the sedecal integrated console went blank.Reboots of the console would not bring up the screen.Patient had to be moved to another room.There was a "foob" for the console.Intermittently generator console (old style) freezes and will not reboot.New console will not power on/off generator.The first replacement console was reported as a failure out of box and returned to lf for evaluation on rga 57221811.Evaluation of this console was done by r and d ((b)(6)) who found that the console would power on but not turn on the generator.Further investigation found cause was a broken c1 capacitor on the `turn on' pcb located at the bottom of the console enclosure.The console was returned to sedecal for repair.
|