• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°; HF-RESECTION ELECTRODES Back to Search Results
Model Number A22251A
Device Problems Detachment Of Device Component (1104); Component Falling (1105)
Patient Problem Foreign Body In Patient (2687)
Event Date 06/07/2017
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device has not yet been returned to the manufacturer for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during an unspecified therapeutic transurethral resection (tur) procedure, the roller at the distal end of the hf resection electrode detached and fell inside the patient.However, no fragment remained inside the patient since the roller was reportedly retrieved.The intended procedure was successfully completed with another similar device and there was no adverse event or patient injury.
 
Manufacturer Narrative
The suspect medical device was returned to the manufacturer for evaluation/investigation.The evaluation/investigation confirmed that the roller at the distal end of the hf resection electrode is detached and that the wire holding the roller is severly worn.This damage is most likely caused by using the monopolar hf resection electrode in bipolar mode.Therefore, this event/incident was attributed to abnormal use/off-label use.Furthermore, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf resection electrode without showing any abnormalities.The case will be closed from olympus side with no further actions.However, the event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results and retrained to correctly use the olympus medical devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer (Section G)
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O.
telickova 457/29
prerov
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 
4940669662
MDR Report Key6666131
MDR Text Key78589766
Report Number9610773-2017-00088
Device Sequence Number1
Product Code FAS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK951863
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 09/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22251A
Device Catalogue NumberA22251A
Device Lot Number15055P01L001
Other Device ID Number04042761021562
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-