Model Number 9226
Device Problem No Audible Alarm (1019)
Patient Problem Low Oxygen Saturation (2477)
Event Date 06/03/2017
Event Type  malfunction  
Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event Description
It was reported: "the customer was asking about the set-up of the rad-87 and reported a patient had coded this weekend.She says the therapist on this weekend was standing by the patient.The patient coded.The therapist was wondering why they didn't hear an alarm.It's unclear whether the monitor was flashing an alarm.The monitor did show hash marks with no data within a minute of the code.The monitor was set on a 15 sec delay.The patient was resuscitated and no complications reported.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-87
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
Manufacturer (Section G)
INDUSTRIAL VALLERA DE MEXICALI S.A DE C.V.
calzada del oro, no. 2001
parque industrial palaco
mexicali, baja california 21600
MX   21600
Manufacturer Contact
stefan lissmann
52 discovery
irvine, CA 92618-1604
9492977000
MDR Report Key6669098
MDR Text Key78468657
Report Number2031172-2017-00679
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091241
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9226
Device Catalogue Number9226
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received06/05/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-