• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. IMPACTOR-POLY LINER-36MM; HIP JOINT METAL/POLYMER/METAL SEMI-CONSTRAINED POROUS-COATED UNCEMENTED PROSTHES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. IMPACTOR-POLY LINER-36MM; HIP JOINT METAL/POLYMER/METAL SEMI-CONSTRAINED POROUS-COATED UNCEMENTED PROSTHES Back to Search Results
Catalog Number 112636
Device Problems Material Deformation (2976); Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/07/2016
Event Type  malfunction  
Manufacturer Narrative
The event is regarding damaged threads involving a mako impactor was reported.The event was confirmed.Method & results: device evaluation and results: material analysis engineer indicated that the damage on threads consistent with cross threading.Medical records received and evaluation: not performed as medical records were not provided for review.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there has been no other event for the lot referenced.Conclusions: the investigation confirmed that the device has damaged on threads.Material analysis engineer indicated that the damage on threads consistent with cross threading.If additional information becomes available, this investigation will be reopened.
 
Event Description
Broach handle the locking pin came out.Liner impactor is stripped and broach handle the locking pin came out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPACTOR-POLY LINER-36MM
Type of Device
HIP JOINT METAL/POLYMER/METAL SEMI-CONSTRAINED POROUS-COATED UNCEMENTED PROSTHES
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6669269
MDR Text Key78463102
Report Number3005985723-2017-00278
Device Sequence Number1
Product Code OQG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number112636
Device Lot Number120124
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2017
Date Manufacturer Received05/31/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/21/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age43 YR
-
-