The event is regarding damaged threads involving a mako impactor was reported.The event was confirmed.Method & results: device evaluation and results: material analysis engineer indicated that the damage on threads consistent with cross threading.Medical records received and evaluation: not performed as medical records were not provided for review.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there has been no other event for the lot referenced.Conclusions: the investigation confirmed that the device has damaged on threads.Material analysis engineer indicated that the damage on threads consistent with cross threading.If additional information becomes available, this investigation will be reopened.
|