BOSTON SCIENTIFIC - GALWAY WATCHMAN® ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
|
Back to Search Results |
|
Model Number M635TU20060 |
Device Problems
Leak/Splash (1354); Difficult to Open or Close (2921)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/16/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.A review of the batch history, historical trending, and similar complaint trending review for the product family will be conducted.If there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
|
|
Event Description
|
It was reported a hemostasis valve leak occurred.A left atrial appendage (laa) closure procedure was being performed.A watchman® access system was used.When the access system crossed to the left side, the wire was removed and the hemostasis valve could not be tightened completely; it came unthreaded.They attempted to open the valve completely and re-close it, but that did not solve the problem.The hemostasis valve was leaking a minimal amount of blood, but there were no complications to the patient.They removed the access system from the patient and inserted an amplatz wire to exchange for a new access system.The procedure was successfully completed with the new access system.
|
|