MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problems Unintended Arm Motion (1033); Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/01/2017
Event Type  malfunction  
Manufacturer Narrative
No parts were received by manufacturer for analysis.
 
Event Description
A medtronic representative reported that during a cranial resection procedure the vertek arm was not tightening.The site switched to another vertek arm and completed the surgery.There was no impact on patient outcome.There was a delay of less than 1 hour.
 
Manufacturer Narrative
Both vertek arms were returned to manufacturer.Analysis of the arms could not confirm any functional issues as both locked and released as intended upon testing, despite being visibly worn and discolored.No problem found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
sam saleh
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key6671246
MDR Text Key78551281
Report Number1723170-2017-02679
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 08/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Device Lot Number120802
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received08/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/18/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age24 YR
Patient Weight113
-
-