ABBOTT MANUFACTURING INC ARCHITECT C8000 SYSTEM; AUTOMATED CHEMISTRY ANALYZER Back to Search Results
Catalog Number 01G06-11
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.An evaluation is in process.
 
Event Description
The customer observed falsely elevated sodium patient results while using the architect c8000 analyzer ict module.The following data was provided.The customer uses normal range 133 to 143, critical greater than 160.Sid (b)(6) initial 161, repeat using two different analyzers 163, 163 inova method was lower, 141, and the advia method was similar, 161.The normal range for these methods was not provided.An additional specimen was drawn at another location and was lower, 143, using the architect method.The historical result for this patient is normally around 140.No impact to patient management was reported.
 
Manufacturer Narrative
Evaluation of the customer issue included a review of the complaint text, a search for similar complaints, a labeling review, instrument log review, and an instrument service review.No returns were made available from the customer site for this evaluation.No adverse trend was identified for the customer issue.Labeling was reviewed and found to be adequate.Service history review identified no contributing factors to the customer issue.Based on all available information and the evaluation, no systemic issue was identified and no product deficiency was identified.
 
Manufacturer Narrative
The device evaluation was reassessed and concluded that no malfunction occurred, the device did perform as intended.On july 24, 2017, the customer informed abbott customer support that the original sample was collected in a sodium citrate tube instead of lithium heparin tube which caused interference with the sodium results for sid 17061207178.Product labeling does not include sodium citrate as an acceptable anticoagulant collection vessel for use with the test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT C8000 SYSTEM
Type of Device
AUTOMATED CHEMISTRY ANALYZER
Manufacturer (Section D)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer (Section G)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key6671740
MDR Text Key78555480
Report Number1628664-2017-00273
Device Sequence Number0
Product Code JGS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K980367
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 08/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2017
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number01G06-11
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/24/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-