An endo stitch suturing device was opened to the sterile field.As the surgeon went to use it during the case, it would not work properly.The device was removed from the field and a new device was successfully used.The procedure was completed as planned and no harm came to the patient.The device was sent to clinical engineering and upon further investigation, it appeared that the control buttons that pull the suture into the device were not evenly pushing down, therefore creating the potential for difficult manipulation by the surgeon.E-mailed photos.The device will be returned to the manufacturer for failure analysis.
|