• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ GMBH & CO. KG TUTTLINGEN, GERMANY BIPOLAR CUTTING LOOP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ GMBH & CO. KG TUTTLINGEN, GERMANY BIPOLAR CUTTING LOOP Back to Search Results
Model Number 26040GP1
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/31/2017
Event Type  malfunction  
Manufacturer Narrative
Evaluation of the instrument confirmed that the wire loop has broken off the distal end; it is possible it was compromised by mechanical force; we cannot confirm.
 
Event Description
Allegedly, during a hysterectomy procedure the doctor noted that the loop broke and fell into the patient.The doctor immediately removed the broken piece and replaced the electrode and went on to complete the case.The hospital reported there was no injury to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOLAR CUTTING LOOP
Type of Device
BIPOLAR CUTTING LOOP
Manufacturer (Section D)
KARL STORZ GMBH & CO. KG TUTTLINGEN, GERMANY
mittlestrasse 8,
78503
tuttlingen, germany,
GM 
Manufacturer (Section G)
KARL STORZ GMBH & CO. KG TUTTLINGEN, GERMANY
mittlestrasse 8,
78503
tuttlingen, germany,
GM  
Manufacturer Contact
susie chen
2151 e. grand avenue
el segundo, CA 90245-5017
4242188519
MDR Report Key6677542
MDR Text Key78711403
Report Number9610617-2017-00051
Device Sequence Number1
Product Code HIN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122983
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 06/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model Number26040GP1
Device Catalogue Number26040GP1
Device Lot Number52435
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received06/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age36 YR
Patient Weight65
-
-