• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET ORTHOPEDICS G7 NEUTRAL ARCOMXL LINER; HIP PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET ORTHOPEDICS G7 NEUTRAL ARCOMXL LINER; HIP PROSTHESIS Back to Search Results
Model Number N/A
Device Problem Insufficient Information (3190)
Patient Problems Pain (1994); Weakness (2145); Ambulation Difficulties (2544)
Event Date 05/09/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: 51-101130 taperloc femoral stem 3608363, 11-363662 modular femoral head 271720, 110017102 g7 finned 3 hole shell 3648662.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it remains implanted in the patient.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports:0001825034-2017-04618, 04628, 04629.
 
Event Description
It was reported that a patient underwent a hip arthroplasty approximately a year and a half ago is experiencing pain, weakness in legs, and abnormal gait.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Review of the complaint history determined that no further actions are required.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL ARCOMXL LINER
Type of Device
HIP PROSTHESIS
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6693727
MDR Text Key79288448
Report Number0001825034-2017-04623
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,study
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date02/12/2020
Device Model NumberN/A
Device Catalogue Number010000739
Device Lot Number3517228
Other Device ID NumberSEE H10 NARRATIVE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/30/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/17/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age89 YR
Patient Weight84
-
-