• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE INJECTOR; INJECTOR AND SHEATHSET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE INJECTOR; INJECTOR AND SHEATHSET Back to Search Results
Model Number NM-400U-0523
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/19/2017
Event Type  malfunction  
Manufacturer Narrative
The subject device referenced in this report has not yet been returned to olympus for evaluation.Therefore the exact cause of the reported event could not be conclusively determined at this time.If additional information or the subject device is received at a later time, this report will be supplemented.
 
Event Description
During a procedure regarding the gastroenterology, the ten nm-400u-0523's were used.It was reported that the needles could be extracted a little bit, but the needles could not be retracted into the sheath for all ten devices.The procedure was canceled.There was no patient injury reported.This report describes the third device.
 
Manufacturer Narrative
This is a supplemental report for mfr report # 8010047-2017-00857 to provide additional information based on the evaluation result of the subject device.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.As a result of the evaluation on july 31, 2017, it was confirmed that the needle of the subject device had been retracted into the sheath, but it could not be extended.The sheath of the subject device was buckled at 205 mm and 735 mm from the distal end.The device history record for the lot indicated no abnormality with the event-related items.Based on the similar cases in the past, the needle might be unable to retract into the sheath due to the large friction between the sheath and the needle after the sheath was buckled.The sheath might be buckled, because excessive load was applied to the sheath when the subject device was inserted into the endoscope, when it was taken out from the sterile package, or when it was checked before use.The instruction manual of the device has already warned as follows: when inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.Insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.Stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE INJECTOR
Type of Device
INJECTOR AND SHEATHSET
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
Manufacturer Contact
katsuaki morita
2951 ishikawa-cho
hachioji-shi, tokyo-to 
426425177
MDR Report Key6700066
MDR Text Key80052242
Report Number8010047-2017-00857
Device Sequence Number1
Product Code FBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK902736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 08/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNM-400U-0523
Device Lot NumberK6Y16
Other Device ID Number04953170382291
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/31/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-