Boston scientific received information that the patient with this right atrial (ra) lead heard their device beeping due to high ra lead impedance measurements of greater than 2500 ohms.There was also noise on the ra channel when the patient moved their arm; however, the patient did not require atrial pacing.During the in-office interrogation, lead impedance measurements were normal.The device was reprogrammed and the patient would continue to be monitored.No adverse patient effects were reported.The lead remains in service.
|