The customer reported via phone call that when they try to enter a bolus in their insulin pump, it does not deliver insulin.Their blood glucose is 294 mg/dl.The customer mentioned that they were in the hospital for a few weeks but it was not diabetes related.They stated that while they were in the hospital, they tried checking their bolus but did not receive any insulin.The customer mentioned that they have had issues downloading their readings at home.They also mentioned that their pump does not alarm and that their insulin pump does not show any alarms in the history.They mention that the down arrow is worn down to the point that the metal is showing and that there is damage to the battery cap.The customer performed the self-test on the insulin pump and it passed.The customer was advised to revert to the back-up plan.The insulin pump was returned for analysis.
|
Findings: pump passed all functional testing, including idle current test, run current test, self test, off no power test, unexpected restart alarm error test, basic occlusion test, occlusion test, prime test, excessive no delivery test, displacement test and rewind test.No excessive no delivery alarm noted.Pump programmed with multiple bolus deliveries and all registered properly in the bolus history noted.No bolus delivery anomaly noted.Pump passed the inches.Pump received with minor scratched display window, cracked case at display window corners, cracked battery tube threads, cracked reservoir tube lip, cracked belt clip slot and keypad overlay worn-out.A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.
|