MAUDE Adverse Event Report: CHATTEM ICY HOT SMART RELIEF TENS THERAPY; TRANSCUTANEOUS NERVE STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHATTEM ICY HOT SMART RELIEF TENS THERAPY; TRANSCUTANEOUS NERVE STIMULATOR Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Local Reaction (2035); Burning Sensation (2146)
Event Date 05/01/2017
Event Type  Injury  
Event Description
Initial information regarding this unsolicited case from united states was received on (b)(6) 2017 from the patient.This case involves a (b)(6) old male patient, who developed series of blisters and large patch of skin which appeared to be burned after receiving smart relief tens therapy (icy hot smart relief tens therapy).Past drugs, medical history and concomitant medications were not reported.On an unknown date, the patient started therapy with smart relief tens therapy at an unspecified frequency (lot/batch number and expiry date: unknown) for pain from arthritis and degenerative disc disease.Patient had used smart relief tens therapy for several years, and it was a great relief for patient's pain from arthritis and degenerative disc disease.About a month ago ((b)(6) 2017), patient had a series of blisters developed under a new replacement pad (latency: unknown).Patient let the blisters heal for about a week, then tried another new pad.Within two days he developed a large patch of skin which appeared to be burned (onset date: 2017 and latency: unknown).At the age of (b)(6), he did not use any medications, and would prefer not to begin now.However patient was in pain when he could not use his tens unit.Action taken: drug withdrawn nos.Corrective treatment: not reported.Event outcome: recovered from application site blister on unknown date and unknown for application site burning.Qa review: all retains are examined for correct packaging, seal, missing components, working control unit, legibility of lot numbers /expiration date, correct label and foreign matter.No anomalies were found that could contribute to this complaint.A thorough investigation ((b)(6), capa protocol and review of ihsr [2014]) has been done to investigate.The likelihood of the ihsr producing burns in some users.This investigation found that it is highly unlikely.For the smart relief to produce tissue trauma.The maximum current and power density calculated for the smart relief is a tiny fraction of the limits in the fda guidance.The complete investigation is located in the device files.Lot information was not provided and device return is not expected at this time; therefore, additional investigation is not possible.Pharmacovigilance comment: sanofi company comment dated 30-06-2017:this case concerns an elderly male patient who experienced application site burning and application site blister whilst being treated with smart relief tens therapy.The causal role of smart relief tens therapy cannot be ruled out, however this case lacks information regarding the condition of the skin at the contact area prior to applying the device leads (e.G., wet, dry, open wounds),concomitant.Medications and medical history for comprehensive assessment of the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICY HOT SMART RELIEF TENS THERAPY
Type of Device
TRANSCUTANEOUS NERVE STIMULATOR
Manufacturer (Section D)
CHATTEM
1715 west 38th street
chattanooga TN 37409
Manufacturer Contact
heather schiappacasse
55 corporate drive
55b-220a
bridgewater, NJ 08807
9089817289
MDR Report Key6705177
MDR Text Key79738965
Report Number1022556-2017-00004
Device Sequence Number1
Product Code NUH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131159
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 06/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/12/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/26/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age66 YR
-
-