ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problems Thermal Decomposition of Device (1071); Power Conditioning Problem (1474)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/16/2017
Event Type  malfunction  
Manufacturer Narrative
Final analysis confirmed that the transmitter would not power on.The ac adapter was missing; the power cord had exposed wire.After opening up the transmitter, multiple circuit components were damaged.A burned mark on the inner casing of the transmitter back casing.
 
Event Description
It was reported that after a lightening strike, burnt marks on the transmitter power adapter-power cord were damaged from the storm/lightning.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6706139
MDR Text Key79788880
Report Number2017865-2017-06372
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Other Device ID Number05414734504799
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received08/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/06/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
-
-