Model Number 23789
Device Problem No Audible Alarm (1019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/10/2016
Event Type  malfunction  
Manufacturer Narrative
The returned device was evaluated.During evaluation, the failure was isolated to a defective instrument board.A service history record review reveals that this unit was in the field for over two (2) years with no previous reported issues prior to this reported event.(b)(6).
 
Event Description
The customer reported "speaker failure".No consequences or impact to patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADICAL-7
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
Manufacturer (Section G)
INDUSTRIAL VALLERA DE MEXICALI S.A DE C.V.
calzada del oro, no. 2001
parque industrial palaco
mexicali, baja california 21600
MX   21600
Manufacturer Contact
stefan lissmann
52 discovery
irvine, CA 92618-1604
9492977000
MDR Report Key6709457
MDR Text Key79894524
Report Number2031172-2017-00708
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K140188
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number23789
Device Catalogue Number9500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received07/10/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-