Catalog Number AR-1588T
Device Problem Difficult to Insert (1316)
Patient Problem Tissue Damage (2104)
Event Date 06/26/2017
Event Type  Injury  
Manufacturer Narrative
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.Device history record review revealed nothing relevant to this event.The contribution of the device to the reported event could not be determined as the device was not returned for evaluation therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
 
Event Description
It was reported that during an acl reconstruction procedure, after drilling the femoral tunnel with the 4mm spade tip guide pin (ar-1595t), the surgeon passed the acl tightrope (lot: 10062782, line 198547) while retrieving the white tension sutures.After the button was flipped, the surgeon was unable to tension the graft into the socket at all.After numerous attempts he cut out the acl tightrope and used another of the same with no complications.An unplanned incision was made on the femoral side to cut out the acl tightrope.The tightrope was cut up and completely removed from the patient and discarded by nursing staff after the case was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACL TIGHTROPE
Type of Device
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath, sr mdr analyst
1370 creekside boulevard
naples, FL 34108-1945
8009337001
MDR Report Key6714675
MDR Text Key80073852
Report Number1220246-2017-00250
Device Sequence Number1
Product Code HTY
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K110213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 07/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/01/2021
Device Catalogue NumberAR-1588T
Device Lot Number10062782
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-