• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number T175
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2017
Event Type  Injury  
Manufacturer Narrative
At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
 
Event Description
Boston scientific received information that this implantable cardioverter defibrillator (icd) exhibited premature battery depletion.The patient was told last year that the device had three year longevity remaining.However, few months ago device had reached elective replacement indicator (eri).The icd was explanted.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the pacing and sensing functions were tested.The device operated appropriately, according to its performance specifications, with no interruptions in therapy output or programmer communication at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY 2
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key6718095
MDR Text Key80179502
Report Number2124215-2017-11741
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 08/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/03/2010
Device Model NumberT175
Other Device ID NumberVITALITY 2 VR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received08/04/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/07/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0147; 1860; T175
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age52 YR
-
-