The foam is inserted in the cassette as a convenience for the lab technicians.Due to the nature of the foam, the foam may detach from the cassette lid.As such sakura had added a warning label to each box as an additional safety factor to further warn the user to ensure the foam is in place before use.Furthermore, it is required by the operating manual, and training provided to the technicians to place the tissue on the foam in order to ensure the presence of the foam.Technicians are aware of this as this information is present in the instructions for use and it is also a subject during required hands on training.
|
On (b)(6) 2017, sakura finetek usa, inc.Was notified that a user site in the (b)(6) reported an incident which occurred on (b)(6)2016 and involved sakura's paraform 13x13 cassette.Site stated that one gastric biopsy containing 3 fragments, all 2 mm, was lost during processing and no tissue was made available for histological examination because the lab technician did not verify presence of the foam inside the cassette during embedding process.Furthermore, the lab technician failed to follow the instruction of placing tissue on the foam, as described in the instructions for use.The endoscopic report was issued, stated that the tissue was lost during processing, therefore, no histological assessment could be made.
|